Sealing test paper was "confirmation request forming, sealing and assembly processes," the need for sealing devices running Identification (OQ) requirements according to ISO 11607-2:2006 Part 2,to obtain the device after installation by running the program using its run whether the evidence is expected to determine the extent of, and simultaneously detect the formation process of the design documentation for the detection and recording of integrated performance parameters to use when sealing equipment, sealing effect can be visualized and sealing quality, highlight its sealing defects,sealing machine comes with running the program for printing identification records, to detect problems
and adjust sealing sealing parameters to meet the normal needs sealing to ensure sealing quality,and archived for future batches of the sealing effect of sealing products Verify live retrospective.
According to the second part of the ISO-11607 quality requirements for sealing, sealing line must meet the following requirements:
l Continuous seal width and complete sealing;
l No passage or opening;
l No perforation or tear;
l No delamination or separation;
l Installation Qualification sealer (IQ);
l Installation Verification means to verify the installation is complete sealing machine sealing performance and sealing temperature sealing performance before its put into use, sealing pressure or the speed of the seal, these parameters should be adjusted according to the
material to be sealed to meet the needs of sealing quality;
l If the temperature, pressure and other parameters deviate from the test paper can be highlighted with a color bar;
l Sealing operation identification (OQ);
l Sealing pressure test whether passage or opening lines appear;
l Sealing test whether the pressure is too high or too low;
l Sealing test whether the temperature is too high or too low;
l Test whether continuous or sealing leaks occur;